OPERATION
Axio Observer
TFT display touchscreen on the Axio Observer 7, 7 materials
ZEISS
12/2016
431004-7244-001
127
5.11.7
Microscope -> XYZ
The display of the
XYZ
page depends on the microscope stage used.
−
Motorized stages (only CAN bus stages directly connected to 7, 7 materials stands):
−
Manual stage:
Z focus drive settings only (no XY controls are available),
Measure
tab is not available
−
Manual stage / manual Z focus drive:
XYZ
page is not available
The system detects whether a motorized stage is installed during microscope initialization. The
stage should therefore only be changed when the microscope is switched off.
The
Microscope -> XYZ
page contains the tabs
Position
,
Measure
and
Definite Focus
.
5.11.7.1
Position
The controls area of the
Position
tab is divided
into three blocks.
If you are not using a motorized stage,
the XY controls are replaced by a
Start
button (see the following section
(2) Measure
).
(1) Current position display / Set zero
The current Z, X and Y positions are displayed in
millimeters (mm).
The Z position displays the height of the nosepiece position. The X and Y positions display the
stage position.
The two
Set Zero
buttons function for XY and Z as follows:
man
Manually
sets the zero point. The current position is defined as the zero point and the display is
set to zero.
auto
Automatically
sets the zero point. The stage / nosepiece moves to the end position which has
been defined as the zero point. The display is then set to zero.
Unit
Selects unit [mm] or [inch]
Fig. 112
Microscope -> XYZ -> Position page