CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axio Observer
Maintenance
ZEISS
12/2016
431004-7244-001
177
6.2.2
Changing the fuses in the
Axio Observer 3, 3 materials, and
5, 5 materials stands
Pull out the power plug before changing
fuses.
The power fuses for the input voltage are located
in the back of the stand of the Axio Observer
3 materials and 5 materials. The fuse compartment
is integrated into the power socket and contains
two T 5.0 A/H 250 V 5 x 20 mm fuses. If the device
fuses fail the cause must first of all be ascertained
and technical problems properly remedied.
•
Pull out the power plug.
•
Pull out the fuse holder (Fig. 164/
2
) frontward.
•
Remove the fuses from the fuse holder and
insert new fuses.
•
Reinsert the fuse holder into the fuse
compartment (Fig. 164/
1
) until it clicks into
place.
•
Plug in the power plug.
Fig. 164
Changing the fuses in the Axio
Observer 3, 3 materials and 5,
5 materials