INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
ZEISS
Fitting the microscope stages
Axio Observer
46
431004-7244-001
12/2016
Y direction:
•
To adjust the front or rear mechanical stops in Y
direction, first loosen the screws (Fig. 24/
1
) at
the covers on the top side of the stage and
remove the covers (Fig. 24/
2
).
•
Then loosen the stop screws (Fig. 24/
3
), move
the stop as necessary and retighten the screws.
•
Then retighten the screws of the covers.
Scanning stage 130 x 85 mot. CAN is
not compatible with the Aqua Stop!
•
After fitting the stage onto the microscope
stand, connect the cables to the XY drive
(Fig. 25/
1
) and to the stand (Fig. 25/
2
).
Fig. 24
Scanning stage 130x85 mot. P, CAN,
top side
Fig. 25
Scanning stage 130x85 mot. P, CAN,
connections on bottom side