INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
ZEISS
Reflector turret
Axio Observer
58
431004-7244-001
12/2016
•
Tilt up the
excitation
half (Fig. 38/
1
), which
now is on top, and remove it from the retaining
elements (Fig. 38/
5b
) of the lower half
(
emission
4
).
•
Remove beam splitter (Fig. 38/
2
) and spring-
loaded frame (Fig. 38/
3
) from the lower half of
the module.
•
Remove the old beam splitter and carefully
place the new one on the spring-loaded frame
(Fig. 38/
3
) with the reflecting side facing up and
place both parts together into the lower half of
the module. Ensure that the catch on the side
of the spring-loaded frame is positioned in the
recess in the lower half of the module.
The reflective (coated) side (Fig. 40/
3
)
of the beam splitter has a beveled
edge (Fig. 40/
1
) or corner (Fig. 40/
2
).
•
Place the
excitation
half of the module
1
) on top of the
emission
half
4
) (retaining elements Fig. 38/
5b
engage with eyelets Fig. 38/
5a
). Holding the
two halves together, turn the module back into
the position for insertion.
•
Reinsert the slotted screws and screw tight.
•
Finally, attach the adhesive label with the name
of the filter combination to the side of the
module.
Fig. 39
Changing the beam splitter
Fig. 40
Labeling on the beam splitter