INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
ZEISS
HAL 100 illuminator
Axio Observer
66
431004-7244-001
12/2016
4.22.2
Adjusting the HAL 100 illuminator
(1)
Coarse adjustment
•
After loosening the clamping screw (Fig. 50/
1
or
8
), remove the halogen illuminator
3
) which is ready for use from the
microscope stand.
•
Switch on the microscope.
•
Direct the light beam to a projection surface
(wall) at a distance of at least 3 m.
Do not look into the light exit aperture
of the illuminator.
•
Using a 3 mm ball-headed screwdriver, turn the
adjusting screw (Fig. 51/
3
) until both images of
the lamp filament appear as sharp as possible
on the projection surface.
•
Then turn the two adjusting screws (Fig. 51/
4
and
5
) until the lamp filament of one image
exactly fills the gaps in the reflected filament
image (Fig. 51/
1
).
(2)
Fine adjustment
•
If necessary, loosen the clamping screw
(Fig. 52/
1
) and remove the HAL 100 from the
transmitted light illumination mount.
•
Manually unscrew the diffusion disk (Fig. 52/
2
)
from the mount (anti-clockwise). Hold the edge
of the diffuser by the projections on the disk
(Fig. 52/
3
).
•
Reattach the HAL 100 and tighten the locking
screw.
•
Remove filters (if inserted) from the optical path
or deactivate them.
•
Focus on the specimen using a
≤
40x objective
and find a free area on the specimen.
•
Remove the eyepiece and center the lamp
filament and its reflection in the pupil image
using the adjusting screws (Fig. 51/
4
and
5
).
•
Turn the adjusting screws (Fig. 51/
3
) until the
illumination of the visible image is as
homogeneous as possible.
•
Once you have completed the adjustment,
remove the HAL 100.
•
Manually screw the diffuser into the adaptor.
•
Reattach the HAL 100 and activate the filters.
Fig. 51
Adjusting the HAL 100 illuminator
Fig. 52
Fitting/removing the diffusion disk