INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
ZEISS
HBO 100 illuminator
Axio Observer
68
431004-7244-001
12/2016
4.23
HBO 100 illuminator
4.23.1
Inserting the HBO 103 W/2 mercury vapor short-arc bulb
For safety reasons, the HBO 100 illuminator and the HBO 103 W/2 mercury vapor short-arc bulb are
packed separately.
The first step in setting up the illuminator is therefore to insert the HBO 103 W/2 bulb into the lamp
housing.
Instructions on inserting or replacing the HBO 103 W/2 bulb can be found in the operating manual
supplied with the illuminator.
An FL attenuator, discrete (manual or motorized) should be fitted into reflected-light
illuminator FL to change the transmission.
4.23.2
Fitting the HBO 100 illuminator
•
Remove the cover from the reflected light
socket (Fig. 54/
3
).
•
Insert the dovetail of the lamp housing
(Fig. 54/
1
) into the reflected light socket
3
) on the rear of the instrument and
tighten the clamping screw (Fig. 54/
2
) using the
3 mm ball-headed screwdriver.
•
Insert the multi-pin plug of the HBO 100
illuminator (Fig. 54/
6
) into the device socket
5
) on the HBO 100 power supply unit
4
) and secure with the coupling ring.
•
Insert the power supply cable (Fig. 54/
8
) into
the power connector (Fig. 54/
7
) on the
HBO 100 power supply unit, then insert the
cable plug into a mains socket.
Fig. 54
Power supply unit (ballast unit) for
HBO 100