OPERATION
Axio Observer
Operating and function controls - description
ZEISS
12/2016
431004-7244-001
85
−
Instrument configuration with sideport 60N right (R 100) and left (L 100); 3 switch positions (425155-
0000-000):
100 % vis: 0 % left/right
0 % vis : 100 % doc left
0 % vis : 100 % doc right
−
Instrument configuration with sideport 60N right (R 100) and left (L 80); 3 switch positions (425165-
0000-000):
100 % vis: 0% doc
0 % vis :80 % doc left
80
0 % vis : 100 % doc right
Nosepiece with objectives (Fig. 57/19)
−
6-position nosepiece H, DIC (Fig. 63/
1
) with DIC
slider slots (Fig. 63/
2
) in all objective positions in
manual, coded or motorized versions
−
Manual and coded nosepieces: rapid objective
change by rotating the nosepiece using the
adjustment ring (Fig. 63/
3
)
−
Motorized nosepiece: rapid objective change
e.g. by using the pair of buttons on the right
control ring (Fig. 77/
1
)
−
The nosepiece can also be operated manually.
If a heating stage or gliding stage is
used, the nosepiece must be lowered
fully using the focus drive before
changing objective - otherwise the
objective may collide with the stage.
Slot for analyzer slider (Fig. 63/4)
−
For fixed analyzer sliders with two
∅
32 mm filter positions, or
±
30° analyzer sliders for DIC (available
for all stands)
Fig. 63
Nosepiece with slot for DIC slider
and PlasDIC slider; analyzer slider
slot