OPERATION
Axio Observer
Operating and function controls - description
ZEISS
12/2016
431004-7244-001
91
Binocular tubes (Fig. 57/27)
−
The binocular tubes provided allow the interpupillary distance to be individually adjusted and the
viewing height to be adjusted within defined limits
Binocular tube 45°/23 with manual shutter vis
−
Switch the shutter on or off by turning the
rotary button (Fig. 71/
1
):
100 % vis
0 % vis (light shutter)
Binocular phototube 45°/23 with sliding prism
for vis / doc, Bertrand lens and manual vis
shutter (Fig. 72)
−
Switch the shutter on and off by sliding the
rotary / sliding button (Fig. 72/
1
):
100 % vis
0 % vis (light shutter)
Bertrand lens
−
The Bertrand lens is focused by rotating the
rotary / sliding button (with the Bertrand lens in
position)
−
Switch the beam path (sliding prism vis / doc)
with the sliding button (Fig. 72/
2
):
0 % vis: 100 % doc
50 % vis: 50 % doc
100 % vis: 0 % doc
The binocular phototube camera port is recommended up to 18 mm camera field (related to
intermediate image) e.g. for an Axiocam with 0.63x camera adapter.
Fig. 71
Binocular tube 45°/23
Fig. 72
Binocular tube 45°/23
B