PUTTING THE INSTRUMENT INTO OPERATION
Axio Vert.A1
Mounting Standard Components
Carl Zeiss
05/2012
431030-7044-001
41
•
Place the microscope stand (Fig. 3-1/
3
) on a
non-vibrating, even, hard and non-flammable
worktop.
•
Remove the plastic cover.
•
Remove the protective caps (Fig. 3-1/
1
and
4
).
The filter slider (Fig. 3-1/
2
) is fixed-
mounted, while appropriate filter
glasses (green filter, attenuation
filter, conversion filter) are not yet
inserted during transport.
3.1.2
Attaching the Binocular
(Photo)Tube
All the binocular tubes listed in the system
overview can be attached to the stand as described
below.
The microscope comes with a factory-mounted
50-mm ergo intermediate piece, to which another
50-mm or 25-mm ergo intermediate piece may be
added. If the ergotube is employed, only ergo
intermediate pieces of up to 25 mm should be
used, however. The utilization of the 50-mm
extension would require optical compromises. For
the installation, follow the same procedure as for
the installation of the tube.
To mount or change the tube, proceed as follows:
•
Use ball-headed screwdriver SW 3 to loosen the
Allen screw (Fig. 3-2/
3
) on the ergo
intermediate piece. If necessary, hold the tube
to be changed and remove it toward the front.
•
Remove the dust cap from the tube to be mounted (Fig. 3-2/
2
) and from the ergo intermediate piece,
as applicable.
•
Insert the binocular tube (Fig. 3-2/
1
) with the dovetail ring into the tube port (Fig. 3-2/
4
) of the ergo
intermediate piece, align it with the ergo intermediate piece and retighten the Allen screw (Fig. 3-2/
3
),
using the ball-headed screwdriver.
Fig. 3-1
Unpacking and Setup
Fig. 3-2
Attaching the Binocular Tube