PUTTING THE INSTRUMENT INTO OPERATION
Carl Zeiss
Mounting Standard Components
Axio Vert.A1
46
431030-7044-001
05/2012
•
Once the stage has been mounted on the
microscope stand, plug in the connection cables
to the XY drive (Fig. 3-10/
1
) and to the PC
2
).
To enable scanning stage 130x85 motP;
CAN to be used on Axio Vert.A1 stands,
the stage must be directly connected to
a PC with the aid of the CAN-USB
converter. The stage will be controlled
via the AxioVision software.
3.1.6.4
Mounting Gliding Stage Z
To mount the gliding stage, follow the same procedure as for the mechanical stage.
•
Before placing the gliding stage on the stand, screw out the three support blocks located on the
underside of the gliding stage.
•
Put spacer washers on the three locating points of the stand.
•
Position the gliding stage on the stand and fasten it from the top with three Allen screws. In so doing,
make sure that the screws also go through the holes of the spacer washers.
When the gliding stage is employed, the objective nosepiece needs to be moved to the lowest
position possible before another objective can be rotated into position. Otherwise, the objective
and the gliding stage may collide with one another.