CARE, FUSE REPLACEMENT AND SERVICE
Axiolab 5
Instrument care
ZEISS
05/2019
430037-7444-001
119
5
CARE, FUSE REPLACEMENT AND SERVICE
5.1
Instrument care
The only care required for the Axiolab 5 is as follows:
•
Switch the device off after each use and put the protective cover on (protects against dust and
moisture).
•
Do not set the instrument up in a humid environment (max. humidity
≤
75%).
•
Cover all open tubes with dust caps.
•
Remove dust and loose dirt on visible optical surfaces using a fine brush, blower brush, cotton wool
bud, optical paper or cotton cloth.
•
Remove water-soluble dirt (coffee, cola, etc.) by breathing on it and wiping with a dust-free cotton
cloth or a moistened cloth. A mild cleaning agent may be added to the water.
•
Remove stubborn, oily or greasy dirt (immersion oils, fingerprints) using cotton wool buds or a dust-
free cotton cloth and optical cleanser L.
This cleaning agent is made of 90 vol% petroleum ether and 10 vol% isopropyl alcohol (IPA). The
components are also known by the following names:
Petroleum ether:
surgical spirit, benzine
Isopropanol alcohol: 2-propyl alcohol,
dimethylcarbinol,
2-hydroxypropane
Clean the optical surfaces with circular movements, starting from the middle and working outward to the
edges. Exert only slight pressure on optical surfaces.
ATTENTION
Do not use acetone to clean the front lens of the Pol condenser.
Please comply with the following guidelines if the microscope is to be used in a hot and humid climate:
•
Keep the instrument in bright, dry and well-aired rooms, with humidity
≤
75%; in particular, sensitive
parts such as objectives and eyepieces should be kept in special dry closets.
Precision optical instruments are always susceptible to mold if they are kept and used under the following
conditions:
−
If the relative humidity > 75% over periods exceeding three days at temperatures b15 °C
and +35 °C.
−
if they are set up in dark rooms without sufficient ventilation.
−
if there are dust and fingerprints on optical surfaces.