Technical data
Dimensions (length x width x height)
Axiolab 5 basic microscope stand
without tube (430037-9011-000) ............................................... approx. 304 mm x 210 mm x 357.5 mm
The other stand models differ slightly in depth and significantly in height, depending on the tube used.
An overview of viewing heights (eyepoint heights) of the various tubes can be found on page 19.
An estimate of the height of the stand with the respective tube can be obtained by adding:
−
10 mm to the viewing height in the lower position of the binocular section for tubes with a fixed
viewing angle
−
10 mm to the viewing height of the upper limit for ergo tubes
Weight
Axiolab 5 microscope stand (depending on version and accessories) ............................. approx. 8 to 20 kg
Ambient conditions
Shipping (in packaging):
Permissible ambient temperature ................................................................................... -40 to +70 °C
Permissible humidity (without condensation) .......................................................... max. 75% at 35 °C
Storage:
Permissible ambient temperature .................................................................................. +10 to +40 °C
Permissible humidity (without condensation) .......................................................... max. 75% at 35 °C
Operation:
Permissible ambient temperature .................................................................................. +10 to +40 °C
Permissible relative humidity (without condensation) .............................................. max. 75% at 35 °C
Highest permitted altitude of use .................................................................................... max. 2000 m
Air pressure ........................................................................................................800 hPa to 1060 hPa
Degree of pollution .......................................................................................................................... 2
Operational specifications
Operational area ................................................................................................................. Closed rooms
Protective class .......................................................................................................................................I
Protection type ................................................................................................................................ IP 20
Electrical safety ............................................................ in accordance with DIN EN 61010-1 (IEC 61010-1)
.............................................................................................
in compliance with CSA and UL regulations
Overvoltage category ............................................................................................................................. II
RFI suppression ....................................................................................... complies with EN 55011 Class B
Noise immunity ...................................................................................... complies with DIN EN 61326/A1
Mains voltage for the Axiolab 5 ........................................................................................... 100 to 240 V
Mains frequency ................................................................................................................... 50 to 60 Hz
Power consumption of the Axiolab 5 ............................................................................................ 100 VA
Fuses in compliance with IEC 127
Axiolab 5 microscope stand ............................................................................. 2 x T 3.15 A/H, 5x20 mm