DESCRIPTION OF THE INSTRUMENT
ZEISS
Control and functional elements on microscope components
Axiolab 5
36
430037-7444-001
05/2019
Binocular photo tube 30°/23 (100:0/0:100)
The light can be directed using a shift knob to
either the eyepieces or the mounted camera.
−
Shift knob (Fig. 2-10/
3
) to front
(eye symbol):
100% light to eyepieces.
−
Shift knob (Fig. 2-10/
3
) to the rear (camera
symbol):
100% light to camera.
−
Push-pull rod (Fig. 2-10/
1
) pushed in:
eyepiece shutter closed.
−
Push-pull rod (Fig. 2-10/
1
) pulled out:
eyepiece shutter opened.
−
Particularly for camera shots with extended
exposure times, it is recommended that
possible residual light incidence through the
eyepiece is prevented either by means of an
eyepiece shutter or eyepiece cover (included
in dust guard set). If neither is available,
remove the eyepiece and attach the supplied
dust cap to the eyepiece barrels!
Fig. 2-10
Binocular photo tube 30°/23 with
toggle graduation 100:0/0:100