OPERATION
ZEISS
Illumination and contrast methods in transmitted light
Axiolab 5
96
430037-7444-001
05/2019
•
Then swivel the upper part of the circular polarizer D (Fig. 4-10/
4
) into the beam path.
•
Rotate the lever of the lambda/4 plate of the circular polarizer D (Fig. 4-10/
3
) until maximum
extinction is achieved (dark-gray field of view) (lever points 45° to the right).
1
Slider 6x20mm with lambda/4 plate
2
Lower section of circular polarizer
3
Lever for rotating the lambda/4 plate
4
Lambda/4 plate in the upper part of the circular polarizer
5
Adjustment slits
Fig. 4-10
Components for circular polarization contrast
•
An (anisotropic) specimen should not be examined until after the above adjustment.
•
Reinsert the specimen to be examined.
The interference colors – which depend on the material, specimen thickness and orientation – of the
specimens appear constant and independent of stage rotation.