Mounting standard components
Axiolab 5
70
430037-7444-001
05/2019
3.1.12
Installing or replacing the LED
modules for reflected light
fluorescence
Axiolab 5 fluorescence stands can accommodate
up to three fluorescence LED modules.
To insert or replace the LED modules, proceed as
follows.
•
Switch off the microscope, remove the power
cord on the microscope and allow it to cool
down at least 15 min.
•
Remove the cover; see also section 3.1.11.
•
Unscrew the module holder (Fig. 3-25/
1
) from
the stand socket.
•
When changing the modules, remove the
module plugs from the connectors
(Fig. 3-26/
1.2
) of the stand.
Fig. 3-26
Changing the LED modules
•
Pull the old LED modules (Fig. 3-26/
1.3
) out of the holding positions of the LED tube (Fig. 3-26/
1.1
).
Fig. 3-25
Removing the module holder