Mounting and centering the low-
power system for the objectives
2.5x/4x
•
If necessary, remove the polarizer or filter holder
from the condenser carrier.
•
Hold the low-power system (
4
) parallel
to the bottom of the condenser carrier
(
3
) and screw the bracket bolt
6
) of the low-power system with the
angled adjustment lever (
7
) into the
front threaded opening to the left below the
condenser carrier (
3
) until it stops.
•
With the adjustment lever (
1
), screw
the stop bolt (
2
) into the back
threaded opening of the condenser barrier
(
3
) to the stop.
•
Swing the low-power system into the beam
path. Make sure that the fixture is securely
engaged.
•
Open the aperture diaphragm and field diaphragm completely.
•
Adjust both adjustment screws (
5
) using two Allen wrenches (AF 1.5) until the field of vision
is optimally illuminated.
NOTE
The low-power system can be used only in combination with the condenser 0.9/1.25.
Fig. 3-29