START-UP
Axiolab 5
Connecting to the power supply
ZEISS
05/2019
430037-7444-001
75
3.2.4
Inserting the modulator disk in the
condenser 0.9 BF Pol
•
Remove the condenser (Fig. 3-30/
1
) from the
condenser carrier (see section 3.1.8).
If the condenser cannot be lowered sufficiently,
e.g. with the low-power system mounted, then
it may be necessary to remove it with the stage
carrier, then lower it to the stop and remove the
condenser.
•
Loosen the clamping screw (
5
) of the
condenser’s dial segment (
3
) with the
Allen wrench (AF 3) and pull out the dial
segment.
•
Slide the modulator disk (
4
) with its
two-pronged forked opening pointing forward
into the condenser. Make sure that the disk
engages in the guide on both inner sides of the
condenser. The guide serves as a stop for the
modulator disk. The pin of the disk’s clamping
screw must slide into the orientation groove of
the condenser.
•
Tighten the disk’s clamping screw with the
Allen wrench (AF 3).
•
Replace the condenser in its carrier (see
section 3.1.8).
3.3
Connecting to the power supply
The power supply of the Axiolab 5 is located at the
back of the instrument in all stand models.
•
Connect the microscope (Fig. 3-31/
1
) to the
power supply via a power cord and mains
socket.
•
The Axiolab 5 can be connected to a line
voltage ranging from 100 to 240 VAC, with
frequencies of 50/60 Hz. The power unit is set
automatically
to the line voltage available.
Fig. 3-30
Modulator disk in condenser
0.9 H Pol
Fig. 3-31
Power supply connector on the
back of the stand