Troubleshooting
Troubleshooting on the Axiolab A microscope is limited to only a few actions:
•
Checking the line cable
•
Checking the set instrument voltage and changing the fuses
•
Checking the illumination and changing the lamp
−
Defective fuse(s)
→
change the fuses according to paragraph (
1
)
−
defective halogen lamp
→
change the lamp according to paragraph (
2
) or (
3
)
(1)
Check the Line Cable and Fuses and
Change Fuses
Make sure to disconnect the instru-
ment from the line first!
•
Check line cable (4-1/
1
) and replace it, if
required.
•
Use 2.0 mm watchmaker’s screwdriver to press
the two springs at the side of the fuse holder
(4-1/
2
) to the inside via the groove and remove
fuse holder.
•
Check fuse inserts (4-1/
3
) and replace defective
fuses with new ones. Make sure to use only
T 0.8 A; 250 V; 5
×
20 mm fuses!
•
Insert fuse holder again until the springs at the
side can be heard to engage.
•
Connect the instrument to the line.
Fig. 4-1
Changing the fuses
Summary of Contents for Axiolab A
Page 1: ...Axiolab A Reflected Light Microscope Operating Manual...
Page 16: ...Carl Zeiss Axiolab A 1 6 06 99 B 40 015 e Fig 1 3 Axiolab A microscope configurations...
Page 18: ...Carl Zeiss Axiolab A 1 8 06 99 B 40 015 e Fig 1 4 Axiolab A accessories 1 of 2...
Page 20: ...Carl Zeiss Axiolab A 1 10 06 99 B 40 015 e Fig 1 4 Axiolab A accessories 2 of 2...
Page 26: ...Carl Zeiss Axiolab A 1 16 06 99 B 40 015 e Fig 1 5 Axiolab A function elements 1 of 2...
Page 27: ...Axiolab A Carl Zeiss B 40 015 e 06 99 1 17 Fig 1 5 Axiolab A function elements 2 of 2 1 1 2...
Page 34: ...Carl Zeiss Axiolab A 1 24 06 99 B 40 015 e...
Page 36: ...Carl Zeiss Axiolab A 2 2 B 40 015 e 06 99...
Page 44: ...Carl Zeiss Axiolab A 3 2 B 40 015 e 06 99...
Page 68: ...Carl Zeiss Axiolab A 4 2 B 40 015 e 06 99...