Axiolab A
Carl Zeiss
B 40-015 e 06/99
4-5
(2)
Changing the 6 V, 30 W Reflected-Light
Halogen Lamp
The following procedure is required to change the
lamp of the 6 V, 30 W reflected-light halogen
illuminator:
•
Switch off the lamp supply; if required, allow
the illuminator to cool down for approx. 15
minutes.
•
Disconnect the line cable from the line.
•
Remove the connection cable of the reflected-
light illuminator from the socket at the rear of
the instrument.
•
Turn the cover (4-2/
3
) to the left and remove it
in backward direction.
•
To facilitate the lamp change, loosen the
hexagonal screw (4-2/
1
) by a few turns and
remove the mount (4-2/
2
) from the reflected-
light illuminator (the tools are listed together
with the accessories)
•
Loosen both fixation screws (4-2/
5
) by turning them to the left and remove the carrier plate (4-2/
4
)
containing the lamp.
•
Remove new 6 V, 30 W halogen lamp from the packaging and insert it into the mount (4-2/
2
); one
tip of the ceramic base must engage in the centering notch of the carrier plate, and the carrier plate
must fit well to the second tip.
☞
Do not touch the lamp bulb with your bare hands; if required, clean the bulb using pure
alcohol
before
switching it on for the first time to prevent dirt from burning in.
•
Attach reflected-light illuminator again in such a way that the halogen lamp is positioned vertically.
Tighten hexagonal screw (4-2/
1
).
•
Attach cover (4-2/
3
) and lock it by turning it to the right.
•
Connect the cable of the reflected-light illuminator to the rear of the microscope stand.
Fig. 4-2
Changing the 6 V, 30 W reflected-
light halogen illuminator
Summary of Contents for Axiolab A
Page 1: ...Axiolab A Reflected Light Microscope Operating Manual...
Page 16: ...Carl Zeiss Axiolab A 1 6 06 99 B 40 015 e Fig 1 3 Axiolab A microscope configurations...
Page 18: ...Carl Zeiss Axiolab A 1 8 06 99 B 40 015 e Fig 1 4 Axiolab A accessories 1 of 2...
Page 20: ...Carl Zeiss Axiolab A 1 10 06 99 B 40 015 e Fig 1 4 Axiolab A accessories 2 of 2...
Page 26: ...Carl Zeiss Axiolab A 1 16 06 99 B 40 015 e Fig 1 5 Axiolab A function elements 1 of 2...
Page 27: ...Axiolab A Carl Zeiss B 40 015 e 06 99 1 17 Fig 1 5 Axiolab A function elements 2 of 2 1 1 2...
Page 34: ...Carl Zeiss Axiolab A 1 24 06 99 B 40 015 e...
Page 36: ...Carl Zeiss Axiolab A 2 2 B 40 015 e 06 99...
Page 44: ...Carl Zeiss Axiolab A 3 2 B 40 015 e 06 99...
Page 68: ...Carl Zeiss Axiolab A 4 2 B 40 015 e 06 99...