START-UP
Axioscope 5/7/Vario
Mounting standard components
ZEISS
03/2018
430035-7344-001
53
3.1.10
Attaching the Pol rotary stage
Two rotary stage models are available. The
mounting procedure is the same for each model.
3.1.10.1
Removing the Pol rotary stage
•
Loosen the screw cap (Fig. 3-16/
8
) of spring pin
with about 3 turns.
•
Push the Pol rotary stage (Fig. 3-16/
5
) to the
front against the spring-loaded pin (Fig. 3-16/
9
),
lift it off the stage carrier (Fig. 3-16/
7
) at the
back and remove it upwards.
•
Retighten the screw cap (Fig. 3-16/
8
).
3.1.10.2
Attaching the Pol rotary stage
•
If necessary, loosen the screw cap (Fig. 3-16/
8
)
of the spring pin with about 3 turns.
•
Place the Pol rotary stage with the groove of the
dovetail (stage bottom) on the spring-loaded
pin (Fig. 3-16/
9
).
•
Attach the rotary stage (432035-9191-000) with the clamping screw (Fig. 3-16/
10
) pointing to the
front right.
•
Push the Pol rotary stage to the front against the spring-loaded pin and lower it at the back into the
stage carrier (Fig. 3-16/
7
), then release it.
•
Retighten the screw cap (Fig. 3-16/
8
).
NOTE
When you are using the rotary stage (432035-9131-000), the click stop must point to the
front.
Fig. 3-16
Changing the Pol rotary stage with
click stop, attachable Pol object
guide and stage clips