START-UP
ZEISS
Mounting standard components
Axioscope 5/7/Vario
64
430035-7344-001
03/2018
3.1.19.2
Adjusting the HAL 100
Coarse setting
•
After loosening the clamping screw (
1
or
9
), remove the ready-to-operate
halogen illuminator (
2
) from the
microscope stand.
•
Switch on the external power supply unit of the
HAL 100 (see section 3.3.3)
431H
.
•
Direct the light beam towards a projecting
surface (wall) which is at least 3 m away.
CAUTION
Never look directly into the light-
emitting aperture of the illuminator!
•
Adjust the adjustment screw (
3
) using
an Allen wrench (AF 3) until both images of the
lamp filament appear as sharp as possible on
the projection surface.
•
Adjust the adjustment screws (
4
and
5
) so that the lamp filament of one image exactly fills
the gaps of the reflected image (
1
).
Fine setting
•
Mount the illuminator on the microscope stand again and tighten the clamping screw to lock it.
•
If applicable, remove the filter from the beam path.
•
Focus the objective
≤
40x on the flat specular surface of a sample and go to a free area of the
specimen.
•
Remove the eyepiece and center the lamp filament and its reflection in the eye pupil image with the
adjustment screws (
4
and
5
).
•
Optimize the even illumination of the pupil image with the adjustment screws (
3
).
•
If applicable, replace the filter in the beam path.
Fig. 3-30
Adjusting the HAL 100