START-UP
ZEISS
Mounting standard components
Axioscope 5/7/Vario
66
430035-7344-001
03/2018
3.1.20
Inserting the adjustment tool of
HBO 100 into the stand Bio-TL/RL
•
Remove the cover (
1
) from the
mounting aperture of the adjustment tool.
•
Insert the adjustment tool (
4
) and
tighten the three included screws (
3
).
•
Attach the cap to the opening (
2
).
Make sure it locks.
•
Slide in the movable connecting piece of the
adjustment tool.
NOTE
The adjustment tool is used to adjust
the HBO 100
illuminator;
see
section 3.1.21.3.
3.1.21
Mounting and adjusting the HBO 100 illuminator
3.1.21.1
Inserting the HBO 103 W/2 mercury vapor short arc lamp
For safety reasons, the HBO 100 illuminator and the HBO 103 W/2 mercury vapor short arc lamp are
packed separately.
NOTE
Please read the manual included with the unit before inserting or replacing the HBO 103 W/2
lamp.
•
Insert the HBO 103 W/2 lamp into the illuminator as described in the accompanying manual.
ATTENTION
Please use an FL attenuator (if it is not installed continuously) to modify the transmission. Gray
filters are not resistant enough on the long run.
Fig. 3-32
Inserting the adjustment tool