START-UP
ZEISS
Mounting optional components
Axioscope 5/7/Vario
80
430035-7344-001
03/2018
3.2.11
Changing the filter in the filter
wheel transmitted light
•
Remove the stage carrier (see section
•
Unscrew three screws (
3
) of the field
diaphragm socket (
4
) and pull the
socket out of the stand base.
•
If necessary, pull the filter to be changed
(
1
) out of the corresponding position
of the filter wheel (
2
).
•
Put the new filter into the position.
NOTE
One filter position is always designed as
firmly installed shutter position.
•
Repeat the procedure for all filter wheel
positions.
•
Put in the field diaphragm socket and screw it in.
•
Mount the stage carrier (see section
Fig. 3-51
Changing the filter in the filter
wheel transmitted light