CARE, FUSE REPLACEMENT AND SERVICE
Axioscope 5/7/Vario
Maintenance and repair work
ZEISS
03/2018
430035-7344-001
131
Problem
Cause
Solution
Halogen lamp 12 V, 50 W flickers,
illumination intensity is not stable.
Halogen lamp 12 V, 50 W is reaching
the end of its life span.
Exchange the halogen lamp 12 V,
50 W,
see p. 60.
Power cable is not installed properly
or is damaged.
Install the power cable properly or
exchange it.
The pins of the halogen lamp 12 V,
50 W are not properly inserted in the
socket.
Insert the pins of the halogen lamp
12 V, 50 W correctly,
see p. 60.
5.4
Maintenance and repair work
Maintenance and repair work on the mechanical, optical and electronic parts inside the instrument and
on the electrical equipment of the Axioscope microscopes may be performed only by the ZEISS Service
Department or specially
authorized
technical personnel.
To ensure that your microscope remains in optimal and faultless working order for the longest possible
time, we recommend that you conclude a service/maintenance contract with ZEISS.
If you want to place an additional order or if a maintenance/repair incident occurs, please contact your
ZEISS representative.
To place a maintenance/repair call, please contact the representative responsible for your area or the
main quarters at
Carl Zeiss Microscopy GmbH
Carl-Zeiss-Promenade 10
07745 Jena, Germany
microscopy@zeiss.com
www.zeiss.com/microscopy
Carl Zeiss Suzhou Co., Ltd.
Modern Industrial Square 3-B, No.333
XingPu Road SIP
215126 Suzhou, China