START-UP
ZEISS
Mounting standard components
Axioscope 5/7/Vario
48
430035-7344-001
03/2018
3.1.7
Mounting the reflector insert
•
First, remove the cover cap (
1
) by
inserting the Allen wrench (AF 3) into the slot
(
2
). Turn the locking screw
counterclockwise and pull the cover cap
forward.
•
Push the reflector insert (
3
) with the
reflector modules P&C (e.g. reflector turret with
6 coded positions) into the upper part of the
stand until it stops, hold it and tighten the
locking screw again.
NOTE
When the reflector insert is removed,
the reflected light path is automatically
blocked so that the glare will not affect
the operator.
3.1.8
Mounting the mechanical stage
Depending on the type, the mechanical stages can
either be stable or they can be rotated and
centered. The shift range is 75 mm in the X-
direction and 50 mm in the Y-direction.
Stages are available with a coaxial gear knob
(
7
) located on the right or the left side of
the stage.
To assemble/disassemble a stage, remove the stage
carrier, then reattach the stage carrier with the
mounted stage to the stand (see section 3.1.15 on
page 59).
(1) Stable mechanical stage, 75x50 R
a) How to remove the stage
•
Remove the four mounting screws (
5
)
on the stage carrier (
6
) using the Allen
wrench (AF 3) (
4
).
•
Remove the stage (
1
) from the stage
carrier. This is easier if you turn the stage and its
carrier over.
Fig. 3-9
Mounting the reflector insert
Fig. 3-10
Changing the mechanical stage,
75x50 R