START-UP
ZEISS
Mounting standard components
Axioscope 5/7/Vario
60
430035-7344-001
03/2018
3.1.16
Changing the 12 V, 50 W halogen
lamp of the HAL 50 halogen
illuminator
•
Switch off the microscope and disconnect it
from the power supply. Wait at least 15
minutes for the halogen lamp to cool down.
•
Remove the HAL 50 halogen illuminator
(
2
) from the back of the stand
1
) and put it down with the open side
facing up.
•
Remove the used 12 V, 50 W halogen lamp (if
necessary) by pulling it out of the cap
(
2
) in an upward motion.
•
Push the new lamp (
1
) with its two
cap pins gently and carefully into the lamp cap
(
2
). Be careful not to touch the
protective glass of the lamp. To avoid bending
its cap pins or damaging the lamp holder, the
lamp must not be tilted.
•
Place the HAL 50 halogen illuminator with the
connecting pins into the back side of the
microscope and push until the lamp engages
securely.
•
Reconnect the microscope to the power supply.
Fig. 3-25
Removing the HAL 50 halogen
illuminator from the stand
Fig. 3-26
Inserting halogen lamp 12 V, 50 W