3.1.19.3
Changing the halogen lamp 12 V,
100 W
CAUTION
Hot surface!
Danger of burns!
Always let the illuminator case cool off
for at least 15 min before touching it!
NOTE
It is not necessary to remove the
illuminator case from the stand when
replacing the halogen lamp.
•
Switch off the external power supply unit of the
HAL 100 (see section
3.3.3). Disconnect the 3-
pin plug from the HAL 100.
•
Press the release button (
3
) of the
halogen illuminator HAL 100 (
1
). Pull
the lamp carrier (
2
) out completely
and put it aside.
•
Press both spring levers
downwards
5
) and pull out the old halogen lamp
6
) in an upward motion.
•
Press both spring levers downwards, insert the
new lamp into the lamp cap (
4
) and
release the spring levers. Use only the included
special tools (
7
) when holding/
grabbing the halogen lamp. Traces of skin
grease on the halogen lamp can impair its life
span.
•
Press the spring levers down again briefly. The lamp is now centered.
•
Replace the lamp carrier and slide it in until it engages fully.
•
Re-connect the 3-pin plug of HAL 100.
ATTENTION
Before switching on the halogen illuminator, remove the tools (
7
) included in the
illuminator case to prevent them from being damaged by heat. The replacement lamp
(
8
) may remain in the illuminator case.
Fig. 3-31
Changing the HAL 100