Mounting the Colibri 3 illumination system and changing the LED modules
3.1.22.1
Mounting the Colibri 3 illumination system on the upper part of the Axioscope
stand
CAUTION
Before removing the Colibri 3 from the microscope stand, please make sure the main power is
switched off. Never look directly into the light-emitting aperture of the lamp.
•
With the Allen wrench (AF 3) (Fig. 3-37/
6
) loosen the clamping screw (Fig. 3-37/
5
) of the RL
illumination connector on the upper back of the stand.
•
Align the illuminator’s (Fig. 3-37/
1
) dovetail ring with the connection piece (
2
) at the stand.
•
Tighten the clamping screw (Fig. 3-37/
5
) with the Allen wrench (AF 3).
•
Connect the illuminator plug (Fig. 3-37/
4
) to the stand socket (Fig. 3-37/
3
).
•
Tighten the fixing screws on the plug.
Fig. 3-37
Mounting the illumination system Colibri 3