OPERATION
Axioscope 5/7/Vario
Default setting of the microscope
ZEISS
03/2018
430035-7344-001
85
4
OPERATION
4.1
Default setting of the microscope
4.1.1
Setting the inter-pupillary distance on the binocular tube
•
Adjust the individual inter-pupillary distance by
symmetrically swiveling of the two eyepiece
tubes towards each other (
The correct inter-pupillary distance has been set
when the observer sees only
one
round image
when looking into the eyepieces!
4.1.2
Setting the viewing height
•
Adjust the viewing height to your individual
requirements by swiveling the eyepiece tubes
up (
A
) or down (
B
).
The binocular ergonomic (photo) tubes (425511-
0000-000, 425512-0000-000, 445514-0000-000
and 425520-9050-000) are equipped with a
continuous height adjustment in the range from
44 mm to 50 mm.
The binocular piece of the ergo tube 425520-
9050-000 can also be pulled out horizontally by
50 mm.
Fig. 4-1
Setting the inter-pupillary distance
on the binocular tube
Fig. 4-2
Setting the viewing height on the
binocular tube