Axioskop 40
START-UP
Axioskop 40 FL
Initial start-up
Carl Zeiss
B 40-810 e 12/01
2-5
2.1.2
Attaching binocular tube or
phototube
All the binocular tubes listed in the system
overview (refer to Section 1.3) can be attached to
the Axioskop
40 / Axioskop 40 FL stand as
described below.
•
Undo hexagonal screw (2-2/
3
) using the SW3
ball-headed screwdriver. Remove dust caps
(2-2/
2
) from the bottom of the tube and the
dovetailed mount (2-2/
5
) on the stand.
•
Hold the binocular tube (2-2/
1
) or the binocular
phototube in a slightly inclined position and
attach it to the stand mount (2-2/
4
). Turn the
binocular tube to the observation position
required and tighten the hexagonal screw using
the ball-headed screwdriver.
CAUTION
Do not hold the tube by the two eyepiece supports when moving it.
2.1.3
Inserting eyepieces and auxiliary
microscope
•
Remove both dust caps (2-3/
1
and
4
) from the
binocular tube.
•
Remove both eyepieces (2-3/
2
) from their cases
and insert them into the binocular tube as far
as they will go.
•
The auxiliary microscope (2-3/
3
), which is used
to view aperture/phase/VAREL/dark-field stops
and to center phase/VAREL/dark-field stops, can
be inserted into one of the tubes instead of an
eyepiece. The adjustable lens enables you to
focus on these stops.
☞
The eyepieces W-PL 10x/23 Br. foc. and PL 10x/23 Br. foc. can be used with the tubes offered
(refer to Section 1.5).
Fig. 2-2
Attaching the binocular tube
Fig. 2-3
Inserting eyepieces
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...