START-UP
Axioskop 40
Carl Zeiss
Initial start-up
Axioskop 40 FL
2-12
B 40-810 e 12/01
2.1.11
Inserting and replacing the 12 V,
35 W halogen lamp
•
Remove the power cable from the microscope.
•
Turn off the microscope and allow it to cool
down for some 15 minutes.
•
Undo the knurled screw (2-12/
2
) and pull up
the lamp housing (2-12/
1
).
•
Hold the lamp carrier (2-13/
3
) by the knob
(2-12/
3
), pull it forward and move guide axis
(2-13/
4
) into the top click-stop position.
•
Swivel lamp carrier (2-13/
3
) backwards, so you
have access to the lamp holder (2-13/
2
).
•
Remove the used lamp (when replacing it) by
pulling it up out of the lamp holder (2-13/
2
)
without tilting it.
•
Push new lamp (2-13/
1
) with its two socket
pins carefully and without force into the lamp
holder (2-13/
2
). Do not touch the front
protective glass on the lamp. The lamp must
not be tilted, otherwise the socket pins may be
bent or the lamp holder damaged.
•
Swivel the lamp carrier forward again (2-13/
3
)
until it makes contact.
•
Then pull lamp carrier (2-13/
3
) from the top
click-stop position and move it to the bottom
position. Take care to ensure that the guide axis
is not tilted so that it does not become
jammed. The springs (2-13/
5
) will center the
lamp.
•
Wiggle the knob slightly to ensure that the
lamp is lodged exactly in the compartment
funnel (2-13/
6
).
•
Replace lamp housing (2-12/
1
) and lock with
knurled screw (2-12/
2
).
•
Reconnect power cable.
Fig. 2-12
Removing the lamp housing
Fig. 2-13
Inserting the 12 V, 35 W halogen
lamp
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...