START-UP
Axioskop 40
Carl Zeiss
Initial start-up
Axioskop 40 FL
2-16
B 40-810 e 12/01
Fine adjustment without adjustment guide
•
Attach the HBO 103 illuminator to the
microscope stand as described in Section 2.1.12
(2).
•
Remove one objective from the nosepiece and
switch the empty opening into the function
position.
•
Place a white sheet of paper onto the
mechanical stage.
•
Completely open the pre-centered luminous-
field and aperture diaphragms in the incident-
light illumination beam path.
•
The pupil image with the two focal spot images
is now visible on the white sheet of paper.
When required, reduce the light intensity using
attenuation filters in the 3-position or
6-position filter slider.
•
Use knurled knob (2-17/
6
) to focus on the brighter focal spot via the collector adjustment.
•
Use the SW 3 ball-headed screwdriver on adjusting screw (2-17/
4
) to focus on the slightly darker,
reflected focal spot of the lamp.
•
If required, set knurled knob (2-17/
6
) and adjusting screw (2-17/
4
) repeatedly until both focal spots
have the same dimension or size.
•
Use adjusting screws (2-17/
1
and
3
) to set the brighter focal spot in accordance with Fig. 2-20.
•
Use adjusting screws (2-17/
2
and
5
) to set the darker focal spot in accordance with Fig. 2-20.
Fig. 2-20
Focal spot image without
adjustment guide on mechanical
stage
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...