START-UP
Axioskop 40
Carl Zeiss
Initial start-up
Axioskop 40 FL
2-18
B 40-810 e 12/01
•
Focus the collector with knurled knob (2-22/
4
), so that the two arcs are focused clearly.
•
Use the adjusting screw (2-22/
1
) to set the burner in an axial direction to the reflection, so that both
arcs appear the same size in the viewing window of the adjustment guide (refer to Fig. 2-22/
6
).
•
Use the adjusting screws for height (2-22/
2
) and side adjustment (2-22/
3
) to center the arc and
reflection parallel to each other in the adjustment circle of the viewing window (2-22/
6
). The arc and
reflection should not be superimposed on each other.
•
After you finish adjusting, focus again with the knurled knob and push the adjustment guide back.