2.2.2
Inserting stops (Ph, Var) and
neutral filter in slider
•
Undo the four slot screws (2-24/
1
) and remove
cover (2-24/
2
).
•
Insert stops (2-24/
3
or
10
) into the left or right
position with the metal facing down in the
slider (2-24/
5
). If two VAREL segment stops are
concerned, orientate these to the left uniform
with the opening. Place one insert each (2-24/
9
)
into the ring gaps.
•
Hold the neutral filter (2-24/
6
) by the sides and
insert into the middle of the slider.
•
Carefully place spring mount (2-24/
7
) over the
neutral filter and press down until it engages.
•
Replace cover (2-24/
2
) and fasten with four slot
screws (2-24/
1
).
•
Place pertinent labels into the gaps on the cover
(2-24/
2
) with the notches at the front.
•
Insert slider (2-24/
5
) into the condenser
(notches at the front, cover on top) and allow it
to engage.
•
Center stops (2-24/
3
and
10
) with two knurled
screws each (2-24/
4
and
8
), so that the image
matches that in Fig. 3-21.
Fig. 2-24
Inserting stops and neutral filter
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...