Axioskop 40
START-UP
Axioskop 40 FL
Attachments and conversions
Carl Zeiss
B 40-810 e 12/01
2-25
•
Clamp fixing screw (2-31/
5
) onto the heat sink.
•
Insert the new lamp (2-31/
1
) inside the heat
sink with the spring lever (2-31/
4
) pressed
down in the lamp holder (2-31/
2
) provided in
the illuminator mount (2-31/
3
). Take care to
ensure that the cable to the heat sink is free at
the side. Release the spring lever. Do not apply
force to the lamp.
•
Fit the illuminator housing (2-30/
1
) onto the
illuminator mount (2-31/
3
) and clamp it in place
with clamping screw (2-30/
2
).
•
Connect the HBO 103 illuminator plug and
power plug to the HBO 100 W transformer.
•
Adjust the HBO 103 illuminator as described in
Section 2.1.12 (3).
Fig. 2-31
Changing the HBO 103 W/2 mercury
vapor short-arc lamp