Adjusting the travel range on the ergo-
drive
On mechanical stages with ergo-drive, the travel
range of
x
and
y
adjustment can be extended by a
max.
15 mm
by axial movement of the drive
knobs.
(4)
Removing and attaching additional
sleeves
The two drive knobs for mechanical stages with
catalog numbers 000000-1067-325, 000000-
1063-835 and 000000-1067-836 are fitted with
additional sleeves, providing an even finer setting
of the specimen position. They can be removed if
faster specimen movement is more important.
•
Firstly, undo the two clamping screws (2-37/
4
)
on the bottom additional sleeve (2-37/
3
) and
remove the latter downwards, then undo the
two clamping screws (2-37/
2
) on the top
additional sleeve (2-37/
1
) and pull it
downwards as well.
•
Mount the additional sleeves back on the drive knobs in reverse order and tighten both clamping
screws.
(5)
Setting the torque of the two drive knobs on the ergo-drive
The ergo-drive is set to an average torque value at the factory. This setting can be changed as follows:
a)
Adjusting the x direction
•
Remove the additional sleeves (2-37/
1
and
3
) from the drive knobs when necessary. Undo the
clamping screws to do so.
•
Push the x drive knob (2-37/
5
) down and the y drive knob (2-37/
7
) up.
•
Hold the x drive knob (2-37/
5
) and turn the light knurled ring (2-37/
6
) above it clockwise (easy
action) or counterclockwise (sluggish action) until the desired torque is achieved.
Fig. 2-37
Setting the ergo-drive
Summary of Contents for Axioskop 40
Page 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Page 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Page 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Page 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Page 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Page 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...