Carl Zeiss
Axiotech and Axiotech
vario
3-34
B 40-020 e 06/99
3.5.3
Attachment of MC 80
DX
microscope camera (large format with 4"
××××
5" large-frame
attachment)
•
Loosen three hexagonal screws ( 3-31/
9
) and
remove dust cap (3-31/
7
).
•
Insert adapter for MC 80
DX
microscope camera
(3-31/
6
) into the phototube
( 3-31/
8
) and
tighten three hexagonal screws (3-31/
9
).
•
Insert P 10
×
projection lens (3-31/
5
).
•
Attach basic body
( 3-31/
3
) until stop
and
tighten knurled ring (3-31/
4
).
•
Attach 4"
×
5" large-frame
attachment
(3-31/
2
) including the matte screen (3-31/
1
).
•
Connect cable of basic body ( 3-31/
3
) to the
exposure control (3-31/
10
).
•
Insert a film.
•
Connect cable (3-31/
11
) to the line.
☞
For detailed information on the
MC 80
DX
, please see operating
manual B
40-036, MC
80
DX
Microscope Camera, from Carl Zeiss.
Fig. 3-31
Attachment of MC 80
DX
microscope camera (large format)