2.4.2
Attachment of intermediate tube
for image projection
The intermediate tube for image projection
permits the projection of transparencies, crossline
micrometers and grain size image series into the
microscope image.
•
Loosen hexagonal screw (2-7/
12
) using SW 3
ball-headed screwdriver (2-7/
8
) and remove the
binocular phototube (2-7/
1
).
•
Use the lid of the tube lens case (459511-0000-
000) to remove the tube lens (2-7/
2
) from the
binocular tube and store the tube lens in the
case.
•
Insert dovetail (underside) of the adapter plate
(2-7/
10
) into the stand port of the Axiotech and
slightly tighten clamping screw (2-7/
12
).
•
Insert fixation screw (2-7/
9
) into the
countersunk hole of the adapter plate and
screw it into the drilled hole (2-7/
11
) on the
stand.
•
Tighten clamping screw (2-7/
12
) and fixation
screw (2-7/
9
) using SW 3 ball-headed
screwdriver (2-7/
8
).
•
Remove dust caps (2-7/
4
and
6
), attach
intermediate tube for image projection (2-7/
3
),
align it and screw it to the adapter plate with
the hexagonal screw (2-7/
7
).
•
Attach binocular phototube or other tubes to the intermediate tube and tighten hexagonal screw
(2-7/
5
) using the SW 3 ball-headed screwdriver(2-7/
8
).
☞
For further information on the intermediate tube for image projection please see the relevant
operating manual B 40-038 from Carl Zeiss.
Fig. 2-7
Attachment of intermediate tube
for image projection
Summary of Contents for Axiotech
Page 1: ...Axiotech and Axiotechvario Materials Microscopes Operating Manual...
Page 14: ...Carl Zeiss Axiotech and Axiotechvario XIV B 40 020 e 06 99...
Page 16: ...Carl Zeiss Axiotech and Axiotechvario 1 2 B 40 020 e 06 99...
Page 112: ...Carl Zeiss Axiotech and Axiotechvario 3 46 B 40 020 e 06 99...
Page 114: ...Carl Zeiss Axiotech and Axiotechvario 4 2 B 40 020 e 06 99...
Page 124: ...Carl Zeiss Axiotech and Axiotechvario A 2 B 40 020 e 06 99...