Connection of Axiotech 100
The Axiotech 100 microscope features an integrated stabilized 12 V 100 W power unit which uses the
two switchable voltage ranges 100 ... 127 V AC or 220 ... 240 V AC.
☞
Before the instrument is switched on for the first time, it is therefore absolutely necessary to
check whether the voltage set at the sliding switch (2-26/
1
) on the back of the stand complies
with the line voltage!
•
If the preset instrument voltage
does not
comply
with the line voltage,
make sure
to
set the sliding switch (2-26/
1
) to the correct
line voltage using the SW 3 screwdriver!
•
Connect line cable with coupling device (2-
26/
3
) to the instrument socket (2-26/
2
) and
make connection to the line via line connector
(2-26/
4
).
•
Switch the 12 V 100 W halogen lamp on and
off again via the on/off switch (2-25/
4
). In the
ON position, the green line control lamp in the
switch and at least one LED on the lamp
voltage display (2-25/
2
) must light up if the 12
V 100 W halogen lamp is switched on.
•
Set the required image brightness via the
reflected-light illuminance control (2-25/
3
).
Fig. 2-26
Connecting the Axiotech 100 to the
line
Summary of Contents for Axiotech
Page 1: ...Axiotech and Axiotechvario Materials Microscopes Operating Manual...
Page 14: ...Carl Zeiss Axiotech and Axiotechvario XIV B 40 020 e 06 99...
Page 16: ...Carl Zeiss Axiotech and Axiotechvario 1 2 B 40 020 e 06 99...
Page 112: ...Carl Zeiss Axiotech and Axiotechvario 3 46 B 40 020 e 06 99...
Page 114: ...Carl Zeiss Axiotech and Axiotechvario 4 2 B 40 020 e 06 99...
Page 124: ...Carl Zeiss Axiotech and Axiotechvario A 2 B 40 020 e 06 99...