Carl Zeiss
Axiotech and Axiotech
vario
2-30
B 40-020 e 06/99
2.13.4
Connection of power supply for 6 V 30 W halogen illuminator
(Axiotech
vario
100)
Before connecting and switching on the 6 V 10 ... 40 W power supply, check whether the
primary power supply instrument voltage complies with the line voltage! If this is not the case,
contact your nearest Carl Zeiss agency immediately.
•
Connect cable (2-35/
5
) of the transmitted-light illuminator (2-35/
4
) to the relevant sockets of the 6 V
10 ... 40 W power supply unit (2-35/
1
).
•
Connect the line cable (2-35/
3
) to the line and switch on the power supply unit via its On/Off switch.
The red bar on the switch indicates the ON position.
•
Set the required illuminance via the control (2-35/
2
) of the power supply unit.
☞
In transmitted-light applications, a reflector slider can be used in the darkfield position to
increase the light intensity by approx. factor 2, since no beam splitter is in the beam path in
that case.
Fig. 2-35
Connection of power supply for 6 V 30 W halogen illuminator