Axiotech and Axiotech
vario
Carl Zeiss
B 40-020 e 06/99
3-13
Version 1:
•
Remove any light/attenuation filters from the beam path, i.e. pull out the filter slider or switch to the
free light path position.
•
Push HD DIC reflector slider into the reflected-light illuminator until the DIC position is active (
DIC
mark visible on the reflector slider).
•
Place a strongly reflecting specimen on the stage and focus on it.
•
Replace protection slider ( 3-12/
7
) with DIC (
λ
) compensator (3-12/
2
) and push it into the reflector
slider until the first stop (i.e. the DIC (
λ
) compensator is not yet inserted in the beam path).
•
Remove protection cap (3-12/
6
) from the reflected-light illuminator and replace it with the DIC- prism
slider with fine adjustment (3-12/
5
).
•
Insert DIC prism slider with fine adjustment (3-12/
5
) until stop.
☞
When pushed in half the way, the DIC prism slider can remain in the instrument even if the DIC
technique is not applied. When changing from DIC to brightfield H or darkfield D, the DIC
prism slider must be pulled out until stop.
Fig. 3-12
Setting of reflected-light interference contrast
Summary of Contents for Axiotech