The Axiotech and Axiotech
vario
microscopes, including original accessories, may only be used for the
microscope techniques described in this manual.
Particular attention must be paid to the following warning notes:
The manufacturer cannot assume any liability for any other applications, possibly also involving
individual modules or single parts. This also applies to all service or repair work which is not
carried out by authorized service personnel. If this regulation is not observed, all the claims
against warranty will be forfeited.
The power plug must be inserted in a socket featuring a grounding (earth) contact. The
grounding effect must not be made ineffective by an extension cable which does not have a
protective ground wire.
If it is established that the protection measures are no longer effective, the instrument must be
switched off and safeguarded against inadvertent operation. Contact a Carl Zeiss service
agency or the Carl Zeiss microscopy service for the repair of the instrument.
The wide-range power unit which is integrated in the stand of the Axiotech 30 microscope
permits the use of line voltages in the range between 100 and 240 V ± 10%, 50 - 60 Hz,
without the need for the voltage to be changed at the instrument.
The time-tried, stabilized 12V 100 W power unit, which operates in the two switchable line
voltage ranges 115 V and 230 V AC ± 10 %, 50-60 Hz, is integrated in the stand of the
Axiotech 100 microscope.
Before the start-up of the Axiotech 100, compare the instrument voltage preset at the 230/115
V AC sliding switch on the rear of the stand to the available line voltage and change it, if
required!
Depending on the illumination equipment, only external power units are used for the
Axiotech
vario
100, which must also be compared and adapted - if required - to the existing line
voltage.
Always disconnect the instrument from the line before opening the instrument and before
changing the fuses.
Make sure to use only fuses of the rated power required. The use of makeshift fuses and the
short-circuiting of the fuse holders are not permitted.
Summary of Contents for Axiotech
Page 1: ...Axiotech and Axiotechvario Materials Microscopes Operating Manual...
Page 14: ...Carl Zeiss Axiotech and Axiotechvario XIV B 40 020 e 06 99...
Page 16: ...Carl Zeiss Axiotech and Axiotechvario 1 2 B 40 020 e 06 99...
Page 112: ...Carl Zeiss Axiotech and Axiotechvario 3 46 B 40 020 e 06 99...
Page 114: ...Carl Zeiss Axiotech and Axiotechvario 4 2 B 40 020 e 06 99...
Page 124: ...Carl Zeiss Axiotech and Axiotechvario A 2 B 40 020 e 06 99...