Swing required objective in the beam path .
•
Focus on the specimen via the coaxial coarse/fine drive.
•
Turn large knurled knob of DIC prism slider until the field center appears as dark as possible.
•
Turn small knurled knob of DIC prism slider until the field of view displays homogeneous brightness.
☞
Slightly move the small and the large knurled knob to obtain more homogeneous illumination.
•
Set optimum contrast (depending on the application performed and the specimen) via the large
knurled knob.
☞
After a change of objectives, the three steps described last must be repeated.
•
To change gray contrasts into color contrast, insert the DIC (
λ
) compensator (3-12/
2
) into the HD DIC
reflector slider until stop.
Version 2:
•
Perform the same steps as described for version 1.
☞
The DIC (
λ
) compensator is not required for the DIC compensation prism slider ( 3-12/
4
). The
protection slider can remain in the HD DIC reflector slider.
•
Vary the color contrast by turning the large knurled knob of the DIC compensation prism slider
(3-12/