Axiotech and Axiotech
vario
Carl Zeiss
B 40-020 e 06/99
3-19
3.4.5.2
Setting of epi-fluorescence using the 3 FL reflector slider
•
Perform steps described in sections 3.2 and 3.3.
•
The 3 FL reflector slider (3-17/
3
) replaces the reflector slider HD or HD DIC.
☞
The reflector sliders HD and HD DIC can only be pulled out if one of the dust caps ( 3-19/
5
or
3-17/
2
) has been removed from the left or right side of the stand. To do this, unscrew, the two
hexagonal screws (3-19/
6
) from one side using the SW 3 ball-headed screwdriver, remove the
dust cap and pull out the reflector slider
very carefully
, since a darkfield component projects
from the reflector slider.
•
Attach removed lateral dust cap again.
•
Move filter slider (3-17/
1
) to free light path position.
•
In the center position of the 3 FL reflector slider, select the required specimen spot in reflected-light
brightfield and refocus on it via the coarse/fine drive, if required.
•
Move the user-specific fluorescence filter combinations in the beam path until the left or right stop
position of the 3 FL reflector slider.
•
Open the aperture diaphragm completely.
Fig. 3-17
Setting of epi-fluorescence using the 3 FL reflector slider
Summary of Contents for Axiotech
Page 1: ...Axiotech and Axiotechvario Materials Microscopes Operating Manual...
Page 14: ...Carl Zeiss Axiotech and Axiotechvario XIV B 40 020 e 06 99...
Page 16: ...Carl Zeiss Axiotech and Axiotechvario 1 2 B 40 020 e 06 99...
Page 112: ...Carl Zeiss Axiotech and Axiotechvario 3 46 B 40 020 e 06 99...
Page 114: ...Carl Zeiss Axiotech and Axiotechvario 4 2 B 40 020 e 06 99...
Page 124: ...Carl Zeiss Axiotech and Axiotechvario A 2 B 40 020 e 06 99...