ZEISS
9 Technical Data and Conformity | 9.1 Performance Data and Specifications
9
Technical Data and Conformity
This chapter contains important technical data as well as information on the conformity.
9
.1 Performance Data and Specifications
The microscope must only be operated in closed rooms. It is recommended to install the micro-
scope in a dimmed room where artificial illumination, sunlight or other light sources cannot inter-
fere with image acquisition. The microscope should not be installed near radiators or windows
with direct sunlight.
Compliance with the installation requirements of the microscope and the availability of the re-
quested supplies is the responsibility of the customer and has to be readily available at the time of
installation.
The microscope must be plugged into a properly installed power socket with protective earth con-
tact using the supplied mains cable. The protective earth connection must not be impaired by the
use of extension cables.
Info
Your ZEISS Sales & Service Partner will provide you with the detailed installation requirements.
Weight and Sizes
Main Components
Length
(mm)
Width (mm)
Height
(mm)
Weight (kg)
Axiovert 5 digital
503
244
505
18.2
Air Conditioning
and Quality
Temperature range for operation with indi-
cated performance (24 h per day, regardless
of whether the microscope is in operation or
switched off)
5 to 40
°C
Relative humidity
< 80 % at 40
°C
Atmospheric pressure / altitude
Pollution degree
800 to 1060 hPa /
≤
2000 m above sea level
2
Mains connection
Nominal AC voltage
Nominal frequency
Max. current
L/N/PE 100 to 240 VAC
±
10 %
50/60 Hz
1.4 A
Rating for microscope stand
Protection class
24 VDC, 5 A
IP20 (IEC 60529)
II
Overvoltage Category
Instruction Manual ZEISS Axiovert 5 digital | en-US | Rev. 1 | 431030-7021-101
45