START-UP
Carl Zeiss
Mounting the Camera
Cell Observer SD
42
423638-7044-001
M60-2-0021 e 01/2009
3.5
Mounting the Camera
The following cameras may be connected to the spinning disc unit:
1.0x camera adapter for:
1.2x camera adapter for:
−
AxioCam MRm,
−
AxioCam HRM,
−
AxioCam HSm
−
Photometrics QuantEM
or other EMCCD camera from Photometrics or
Hamamatsu
The procedure for installation is identical for both camera adapter types.
•
Release the two fixing screws (Fig. 3-4/
2
) at the
camera adapter then take the adapter piece
(Fig. 3-4/
1
) out.
•
Remove anti-dust caps from adapter end piece
and camera.
•
Screw the adapter piece (Fig. 3-5/
1
) onto the
2
).
•
Restore the adapter end piece to the camera
adapter. Make sure that the camera is correctly
oriented as you do this (horizontally, with
inscription at the top).
•
Tighten the two screws (Fig. 3-4/
2
) at the
camera adapter.
Fig. 3-4
Removal of adapter piece
Fig. 3-5
Screwing the adapter piece to the
camera