Laser Safety Check
Check the laser safety device for proper function with a laser connected.
1.
Switch on the microscope.
2.
Switch on the laser.
3.
Check the following action items:
Turn reflector turret out of locked
position
Laser shutter closes and laser light point on opaque laser
protection cover extinguishes.
Turn reflector turret into locked position
Laser shutter opens and laser light point appears on
opaque laser protection cover.
Swing transmitted light arm out of
working position
Laser shutter closes and laser light point on opaque laser
protection cover extinguishes.
Swing transmitted light arm into working
position
Laser shutter opens and laser light point appears on
opaque laser protection cover.
You are requested to contact Carl Zeiss MicroImaging GmbH if you intend to use a laser with
wavelengths that differ from those quoted in section 2.7 Technical Data.
If the safety device is found to function in a non-conforming manner, you should shut the
system down, mark it clearly as defective and contact the Service Department of Carl Zeiss
MicroImaging GmbH immediately.