OPERATION
Cell Observer SD
Switching the System off
Carl Zeiss
M60-2-0021 e 01/2009
423638-7044-001
55
4.3
Switching the System off
4.3.1
Turning the Cell Observer SD Off
•
Press the
Shutter
key (Fig. 4-1/
1
) or use the related Axio Vision software tool to close the shutter.
•
Turn the laser off (refer to sections 4.3.2 and 4.3.3).
•
Turn the key switch (Fig. 4-1/
2
) at the spinning disc unit into position
OFF
.
The control LED (Fig. 4-1/
3
) extinguishes.
•
Transfer the main power switch (Fig. 4-4/
1
) of the CSU-X1 control box into Off position.
•
Turn the microscope off.
4.3.2
Turning the Multiline Argon Laser Off
•
Set the laser power controller to minimum level on completion of a working session (potentiometer
(Fig. 4-5/
2
) in left stop position).
•
Transfer the laser power controller into
Standby
mode by flipping the toggle switch (Fig. 4-5/
3
)
down.
•
Turn the key switch (Fig. 4-6/
2
) into position
0
to turn the laser off. The fan will continue to run for
some minutes until the laser head cooled down.
Once the fan has switched off, you can turn the On/Off switch (Fig. 4-6/
3
) at the power supply
unit off.
4.3.3
Turning the Multi Laser Module Off
•
Turn the key switch (Fig. 4-8/
1
) of the laser controller into position
OFF
.
This will shut all laser lines down. The red control LED’s extinguish.
•
Terminate your AxioVision software session.
•
Toggle the On/Off switch into position
0
to turn the signal distribution box off (Fig. 4-7/
1
).
The green control LED (built into On/Off switch) extinguishes.
•
Turn the power strip off. This will cut power supply to the multi laser module.