Instructions for Use
5 Operation
CLARUS 500
5.5 Acquire Images
2660021165005 Rev. A 2017-08
75 / 144
u
Repeat
Next scan
steps for each scan in the series (2-6 total).
5.5.8 Capturing a Stereo Image
A stereo image is comprised of two separate images of the same
eye, offset to produce a stereoscopic pair.
A pair of images of the same eye created as a stereo pair are stored
the pair using
Stereometric Relationship
in DICOM. In the
Re-
view
screen, stereoscopic pairs display as linked viewports for si-
multaneous pan and zoom.
Prerequisite
þ
The device is prepared. (See Preparing the Device [
.)
þ
The patient is prepared. (See Preparing the Patient [
Action
u
Under
Scans
, check
Stereo
.
u
Select the Scan Type (
Color
,
IR
,
FAF-Green
, or
FAF-Blue
). For
more information about Scan Types, refer to: Scan Types
[
.
u
Under
Settings
, select
Myd
if the patient's pupil is dilated and
Non Myd
if the patient's pupil is not dilated.
u
Adjust the alignment for the first stereo image (1) (offset to one
side of the pupil).
1
2
u
Instruct the patient to look inside the imaging aperture, focus
on the green fixation target and blink naturally. If the patient
cannot see the fixation target, have the patient focus on the
external fixation target (Using the External Fixation Target
[
).
Summary of Contents for CLARUS 500
Page 1: ...CLARUS 500 Instructions for Use ...
Page 98: ...Empty page for your notes ...
Page 122: ...Empty page for your notes ...
Page 124: ...Empty page for your notes ...
Page 130: ...Empty page for your notes ...
Page 132: ...Empty page for your notes ...
Page 134: ...Empty page for your notes ...
Page 136: ...Empty page for your notes ...