10 Technical Data and Conformity | 10.2 Installation Requirements
ZEISS
Parameter
Description
Pneumatically retractable multi-mode annular Scanning Transmission
Electron Microscopy (aSTEM) detector with 12 x sample holder; en-
ables bright field (BF), dark field (DF), oriented darkfield (ODF) and
high angle annular dark field (HAADF) transmission imaging
Available in two configurations: 1 channel read out (aSTEM1) or 4
channel parallel read out (aSTEM4)
aBSD detector (optional):
Pneumatically retractable 5 or 6 segment multi-mode solid state BSE
detector; enables materials contrast, crystal orientation, and topo-
graphic imaging
Available in two configurations: 1 channel read out (aBSD1) or 4
channel parallel read out (aBSD4)
Chamber camera
Color CCD camera
with white-light illumination and IR illumination
Specimen current
monitor
6 range auto ranging for precise current measurement in the range of
1 pA to 10 µA
For more details, refer to the document Product Specification.
10.2 Installation Requirements
For a complete list of the installation requirements, refer to the document Installation Require-
ments.
Location
Requirements
Parameter
Requirement
Installation site
Exclusively inside buildings
Recommended
room size
Min. 4.0 m × 5.0 m × 2.3 m
Service area
Min. 1.0 m at each side
Entrance
Min. 1.1 m wide
Hallways
Min. 1.3 m wide
Corners
Min. 1.5 m wide
Transport ways
Free of staircases
Installation cate-
gory
II
Floor stability
> 1000 kg/m
2
Exhaust Line
An exhaust line is required to remove the waste gas of the pre-vacuum pump and to transmit it to
the outside.